DN-70 – Mini.

  • Miniature indicator , Modbus Protocol Basic Installations
  • Available sensor : Load cell, torque sensor etc.
  • simple and quick using four push buttons only.
  • High resolving power (24bit A/D)
  • Quick sampling speed (100times/sec)
  • Watch dog
  • Small size(48*48*128mm) / Cutting size : 45.5*45.5mm
  • Weight : about 280g
  • Power Input 24V
  • Option: OP-03: RS485 , OP-06 :Analog out(DAC)

 

SPECIFICATIONS

SpecificationsAccuracyApplication sensorSensor authorized voltage
(the number of Loadcell connection)Zero adjustment rangeInput signalAD converterDA converterDisplayDisplay speedCharacter heightDecimal pointSampling speedZero variationSensitivity variationVoltage output(OPTION)Usage temperature rangeHumidityPower source usedRelay output

Strain guage sensor (350Ω or 700Ω)
DC 5V, Loadcell(350Ω) available 4
100% F·S (Auto Zero)
1 ~ 3.0mV/V
24bit
16bit
-19999 ~ +99999
Appox. 32times/sec
7Segment 8mm
Random setting available
Max 100 times/sec
within 0.5μV/℃
50ppm/℃
DC 0~±10V  , 4~20mA     (Selective use)
-10℃ ~ 60℃
less than 80% RH (no dew condensation)
DC 24V 0.1A
Contact capacity : AC 250V 3A
Relay life time : more than 1,000,000 times
9c130eb171c978a9c9fea0555d58e74e.jpg

DN10W

  • The product have designed specifically to operate with strain gauge sensors such as Fast- measurement of Load cell, Pressure Transducer, LVDT, etc. The setup sequence is simple and quick using four push buttons only.
  • High resolving power (16bit A/D)
  • High presision (more than 0.012% F.S) and stability
  • Quick sampling speed (200times/sec)
  • Watch dog
  • Small size(96*48*128mm) / Cutting size : 92*45mm
  • Weight : about 500g
  • Option: Power Input 24V in accordance with     customer request

DIMEMSIONS

SPECIFICATIONS

Specifications Accuracy
DN10W
Application sensor Strain guage sensor (350Ω or 700Ω)
Sensor authorized voltage
(the number of Loadcell connection)
DC 5V, Loadcell(350Ω) available 4
Zero adjustment range 100% F·S (Auto Zero)
Input signal 0.5 ~ 3mV/V
Display -19999~99999
Display speed Appox 30times/sec(Adjustment available by buff setting)
Character height 7Segment 5 digit LED 14mm
Decimal point Random setting available
A/D converter 16bit, Max. 100 times/sec
D/A converter 12bit
Zero variation within 0.5μV/℃
Sensitivity variation within0.01%/℃
Voltage output DC 0~±10V  or 4~20mA     (Selective use)
Usage temperature range -10℃ ~ 60℃
Humidity less than 80% RH (no dew condensation)
Power source used AC 85 ~ 265V 50/60Hz
Relay output ㆍUpper and lower limit comparison output
ㆍ4 relay output
ㆍContact capacity : AC 250V 5A
ㆍRelay life time : more than 5,000,000times
Network RS-232C
Option ㆍBCD Paralle out (OP-01)
ㆍRS485 (OP-02)
ㆍDAC ANALOG OUT (OP-03)
2f2bae658728d08cd2e4a9e3044760ba.jpg

DN20W

  • DN20W have designed specifically to operate with Voltage or Current output sensors such as Pressure Transducer,potentionmeter etc. The setup sequence is simple and quick using four push buttons only.
  • High resolving power (16bit A/D)
  • High presision (more than 0.012% F.S) and stability
  • Quick sampling speed (100times/sec)
  • Built-in Shunt calibration switch
  • Watch dog
  • Small size(96*48*128mm) / Cutting size : 92*45mm

SPECIFICATIONS

Specifications Accuracy
Application sensor Pressure sensor, Linear Pontentiometer(LPS, CTP)
Sensor authorized voltage
(the number of Loadcell connection)
DC 10V, 24V (Dip switch)
Zero adjustment range 100% F·S (Auto Zero)
Input signal DC 0~5V, 0~10V, 4~20mA
Display -19999~99999
Display speed Appox 30times/sec(Adjustment available by buff setting)
Character height 7Segment 5 digit LED 14mm
Decimal point Random setting available
A/D converter 16bit, Max. 100 times/sec
D/A converter 12bit
Zero variation within 0.5μV/℃
Sensitivity variation within0.01%/℃
Voltage output DC 0~10V or 4~20mA
Usage temperature range -10℃ ~ 60℃
Humidity less than 80% RH (no dew condensation)
Power source used 220V 50/60Hz (110V available by internal operation)
Relay output AC 250V 5A, Relay life time : more than 5,000,000 times
Network RS-232C
Option ㆍBCD Paralle out (OP-01)
ㆍRS485 (OP-02)
ㆍDAC ANALOG OUT (OP-03)

Indicator

Dacell is providing ultra-precision load cells and various weighting and measuring products,
which can be used in various fields of industries, based on 20 years of experiences in manufacturing
of weighing and measuring products and with own superior technologies.

df8ecba0d8662a9cd2425d550cbe7e24.jpg

DN25W

The product shows signal of differential trane displace sensor (LVDT) into digital display and analogue output as indicator, and intermediate or high sampling (100 times/sec) and comparison output are available
Feature
Best for measuring transformer type displacement sensor (LVDT)
Rapid sampling speed (100 times/sec.)
High resolving power (16bit A/D)
High accuracy (over 0.02% F S) and stability
One touch Auto zero (Display & Analog out zero)
Calibration of measurement Standard
Watch dog
Peak & sample hold
Upper and lower limit comparison output
NG, OK decision function
4 Relay (HH, HI, LO, LL) out
Analog out (DAC)
Option
OP-02 : RS-232C serial interface   OP-03 : RS485 Multi drop interfaceSPECIFICATION

Specifications Accuracy
Application sensor linear variable differential transformer (LVDT)
Sensor authorized voltage 2Vrms/5kHz
Zero adjustment range 100%F·S (Auto Zero)
Display -19999 ~ -99999
Character height 7 segment LED 10mm
Decimal point Random setting available
A/D converter 16bit, 100 times/sec
D/A converter 12bit
Zero variation within 0.5μV/℃
Sensitivity variation within 0.01%/℃
Voltage output DC 0~±10V  or 4~20mA
Usage temperature range -10℃~60℃
Humidity Less than 80% RH (no dew condensation)
Contact capacity AC 250V 5A
Relay life time more than 5,000,000 times
Power source used 220V 50/60Hz (110V available by internal operation)
Option OP-02 : RS232C, OP-03 : RS485
DN-100200-2.png

DN100

  • Suitable for use of sensors measured using strain gauges.
  • Functions such as digital display, analog output, relay output, etc.
  • Suitable for use in places where high-speed measurement is required (pressure load management, impact test, etc.).
  • High-speed responsiveness and high precision.
  • The upper limit setting is displayed.
  •   FEATURE

– Ultra-fast conversion A/D (7,000 sps)

– High resolution (24 bit ADC, 16 bit DAC)

– High precision (0.01% F.S. or higher) and stability

– Actual load and sensor output calibration

– Unit conversion function (kg, N, ℓb, kg/㎠, bar, MPa)

– Auto zero function (Digital display and analog output simultaneously Zero)

FUNCTION

– Peak & sample hold, Average hold

– Upper and lower limits comparison output

– Analog output (DAC)

– NG, OK, judging function

– 4 Relay Contact Output

– Modbus RTU protocol

OPTION

– OP-01 : BCD parallel out   – OP-02 : RS232C serial interface      – OP-03 : RS485 Multi drop interface

– OP-04 : RS422A serial interface     – OP-10: Power DC24V 0.3A     – OP-14: Signal input 5 mV/V

– OP-15: Signal input 10mV/V

Specifications

Application sensor  Strain guage sensors
 Sensor authorized voltage  DC 5V, Loadcell(350Ω) available 4
 Zero adjustment range  100% F.S (Auto Zero)
 Input signal  0.5 ~ 3mV/V
Display  -19999 ~ +99999
 Character height  7 segment  LED 14mm & 10mm
A/D converter  24 bit 7,000 sps
D/A converter  16 bit 7kHz
 Temperature characteristics  ±10ppm/℃
 Analog output  DC 0 ~ ±10V or 4 ~ 20mA (selective use)
 Contact capacity  AC 250V/DC 30V 3A
Relay life time  more than 10,000,000 times
 Usage temperature range  -10℃ ~ 60℃
 Humidity  Less than 80% RH (no dew condensation)
 Power source used  AC 90 ~ 240V, 50/60Hz 12VA
 Sizes  98 × 98 × 141mm
 Panel cutting size  92.5 × 92.5mm
weight  Appr. 700 g

제품치수 (unit:mm)

3973411c2c1ee59e1fe12579a60ef8e6.jpg

DN50W

  • The DN-50W has designed specifically to operate with strain gauge sensors such as Fast – measurement of Loadcell, Pressure Transducer, LVDT, etc. The setup sequence is simple and quick using four push buttons only.
  • And comparison output are available.
  • Sampling speed : Appir. 1000Times Per Second
  • Option: Power Input 24V in accordance with customer request

Feature

  • High resolution (24bit ADC, 16bit DAC)
  • High precision (more than 0.01% F.S) and stability
  • Quick sampling speed (1000times/sec)
  • Relay output : Hi, Low, NG, OK.
  • UNIT CONVERSION : kg, N, ℓb, kg/㎠, bar, MPa
  • Display part and analogue zero by auto zero
  • Analog Peak Hold
  • Small Size (96×48×128mm)
  • Built-in Shunt calibration switch
  • Peak & sample hold
  • Watch dog

OPTION

  • OP-01 : BCD paralle out
  • OP-02 : RS232C serial interface
  • OP-03 : RS485 Multi drop interface

SPECIFICATIONS

Specifications Accuracy
Application sensor Strain guage sensor (350Ω or 700Ω)
Sensor authorized voltage
(the number of Loadcell connection)
DC 5V, Loadcell(350Ω) available 4
Zero adjustment range 100% F·S (Auto Zero)
Input signal 0.5 ~ 3.0mV/V
AD converter 24bit 1000times/sec
DA converter 16bit 1000 times/sec
Display -19999 ~ +99999
Display speed Appox. 32times/sec
Character height 7Segment 14mm
Decimal point Random setting available
Sampling speed Max 1000 times/sec
Zero variation within 0.5μV/℃
Sensitivity variation 50ppm/℃
Voltage output DC 0~±10V  or 4~20mA     (Selective use)
Usage temperature range -10℃ ~ 60℃
Humidity less than 80% RH (no dew condensation)
Power source used AC 85 ~ 265V 50/60Hz
Relay output Contact capacity : AC 250V 5A
Relay life time : over 107  times,  10 times at rated load

DIMENSIONS

DN-100200-2.png

DN 200

  • Suitable for use of sensors measured using strain gauges.
  • Functions such as digital display, analog output, relay output, etc.
  • Suitable for use in places where high-speed measurement is required (pressure load management, impact test, etc.).
  • High-speed responsiveness and high precision.
  • The upper limit setting is displayed.

FEATURE

– Ultra-fast conversion A/D (2,000 sps)

– High resolution (24 bit ADC, 16 bit DAC)

– High precision (0.01% F.S. or higher) and stability

– Actual load and sensor output calibration

– Unit conversion function (kg, N, ℓb, kg/㎠, bar, MPa)

– Auto zero function (Digital display and analog output simultaneously Zero)

FUNCTION

– Peak & sample hold, Average hold

– Upper and lower limits comparison output

– Analog output (DAC)

– NG, OK, judging function

– 4 Relay Contact Output

– Modbus RTU protocol

OPTION

– OP-01 : BCD parallel out                            – OP-02 : RS232C serial interface

– OP-03 : RS485 Multi drop interface       – OP-04 : RS422A serial interface

– OP-10: Power DC24V 0.3A                      – OP-14: Signal input 5 mV/V

– OP-15: Signal input 10mV/V

Specifications.

Application sensor  Strain guage sensors
 Sensor authorized voltage  DC 5V, Loadcell(350Ω) available 4
 Zero adjustment range  100% F.S (Auto Zero)
 Input signal  0.5 ~ 3mV/V
Display  -19999 ~ +99999
 Character height  7 segment  LED 14mm & 10mm
A/D converter  24 bit 2,000 sps
D/A converter  16 bit 2kHz
 Temperature characteristics  ±10ppm/℃
 Analog output  DC 0 ~ ±10V or 4 ~ 20mA (selective use)
 Contact capacity  AC 250V/DC 30V 3A
Relay life time  more than 10,000,000 times
 Usage temperature range  -10℃ ~ 60℃
 Humidity  Less than 80% RH (no dew condensation)
 Power source used  AC 90 ~ 240V, 50/60Hz 12VA
 Sizes  98 × 98 × 141mm
 Panel cutting size  92.5 × 92.5mm
weight  Appr. 700 g

Dimensions

dn130l-p.jpg

DN-130L

  • DN-130L is 2channels indicator to operate with strain gauge sensors such as Fast – measurement of Load cell
  • Pressure Transducer, LVDT, etc. Digital display and analog output, comparison output are available.

Feature

  • High resolution : 24bit ADC, 16bit DAC
  • 2 channel indicator
  • High precision (more than 0.01% FS )and stability
  • UNIT CONVERSION : kg, N, ℓb, kg/㎠, bar, MPa
  • Display part and analogue zero by auto zero (DAC)
  • Watch dog
  • Calibration of actual load and sensor output value
  • Peak & sample hold
  • Relay output : Hi, Low
  • Analog output (DAC)
  • Modbus RTU protocol

Function

– OP-01 : BCD parallel out (chnnel 1)
– OP-02 : RS232C serial interface (Standard)
– OP-03 : RS485 Multi drop interface
– OP-10 : Power DC24V 0.3A
– OP-14 : Input 5mV/V
– OP-15 : Input 10mV/V

SPECIFICATIONS

Specifications Accuracy
Application sensor Strain gauge sensor
Sensor authorized voltage DC 5V
Zero adjustment range 100%FㆍS (Auto Zero)
Input signal 0.5 ~ 3mV/
Display -19999 ~ +99999
Character height 7 Segment LED 14mm
A/D converter 24 bit 100 sps/channel
D/A converter 16 bit 100 Hz/channel
Zero variation within 0.5㎶/℃
Sensitivity variation within 0.01%/℃
Voltage output 0 ~ ±10V or 4~20mA  (Selective use)
Cintact capacity AC 250V/DC 30V 3A
Relay life time More than 10,000,000 times
Isage temperature range -10℃ ~ 50℃
Humidity Less than 80% RH (No dew condensation)
Power source used AC 85 ~ 265V 50/60Hz
Dimensions 98 X 98 X 141mm
Cutting size 92.5 X 92.5mm
Weight About 700g

DN-300

Show totals separately with 2ch indicator.

Left and right values and totals are displayed separately

Analog and contact outputs for each channel and sum individually.

Calibration method by actual load and calibration with rated output of sensor.

Watch dog

Display part and analogue zero by auto zero (DAC)

Analog out(DAC)

Relay output : Hi, Low, NG, OK.

Suitable for use with model RTB15

Option

OP-02 : RS232C serial interface (Standard installed)

OP-03 : RS485 Multi drop interface

 

 

Spec.     model DN-300
SIGNAL 1 ~ 3mV/V
ANALOG OUT DC 0 ~ ±10V, (4 ~ 20mA option)
EXCITATION 5V
Available Sensors Strain gauge type sensor (bridge 120W, 350W, 700W)
Max. Indication -19999 ~ +99999
A/D Converter 24bit 1000 times/sec
D/A Converter 16 bit
Temp. ZERO 0.5μV/℃
SPAN 0.01%/℃
Indicator 7 segment  LED 14mm / 10mm
Status LED Red LED 6 EA
Key switch 4EA
Communication RS232C (A + B)
Comparison Output Hi, Lo
Contact Capacity AC 250V 5A
Temperature -10℃~ 60, 80% RH or less (no condensation)
Weight About 600g
Power supply AC 90 ~ 240V, 50/60㎐
Outer Diameter 97 × 97 × 141 mm (W × H × D)
Panel Cutting Size 92.5 × 92.5 mm

Dimensions

MI4000

Model MI4000 can be applied sensor(4ea) based on Input 4ch, output 4ch. One touch Auto zero.

External data Auto zero function.

External data hold function.

External data peak hold function.

External sample hold function.

External full proof average function.

External peak to peak function.

External program lock function.

Application sensor

Torque transducer, Load cell, Pressure Transducer, Potentiometer, encoder, RPM sensor, L.V.D.T etc.

Option

Analogue Auto zero.

Analogue Scale zero.

SPECIFICATIONS
Temperature storage -10℃ ~ 85℃
Temperature operating 0 ~ 60℃
Relat output AC 125V 0.5A, DC 24V 1A
Transducer excitation 5, 10, 15, 24VDC, ±15VDC
Type of inputs accepted 0.5 ~ 3.5mV/V
Transducer current drive 120mA
Calibration method Front panel switches
Frequency response 5KHz
Full scale output 0~5V, 10V
Low pass filter 10Hz, 100Hz, 1KHz
Temperature drift 0.02% FS/˚C
A/D conversion 16Bit
Sampling speed 500times/sec
Characters displayed 5 digits
Typical accuracy ±0.02%
Polarity indicator Yes
Decimal point selection Push button setup
Display size 14mm & 10mm
Resolution maximum 99999
Type 5X7 DOT 20Digit 4Line
Input/output connector Push button
Weight 1000g
Mounting Panel mounting
Size 96mm × 96mm × 166mm
Power requirements 220V : 50/60Hz(Option 110VAC, DC24V)

« DN-340

pdn10-t.jpg

Pocket indicator PDN10

  • Pocket indicator with battery.
  • The PDN-10 has been designed specifically to operate with strain gauge sensors such as Load cell, Pressure Transducer. LVDT. etc.
  • Peak hold
  • Auto Zero
  • Measuring Data Saving (Max 500pcs.)
  • Hi, Low Display LED

SPECIFICATIONS

Specifications Accuracy
Power Source DC 9V battery
Electric current Consumption 15mA
LOAD CELL input range 1.0mV/V ~ 3.0mV/V, Potenticmeter : 100~10KΩ
Display 4 1/2 Digit LCD
Characters displayed +19999
Display Speed Approx. 10 times/sec
Decunak point Position 0, 0.0, 0.00, 0.000
A/D Sampling speed 100 times/sec
A/D Resolution 16 bit
Zero Adjust Range 100% F.S(AUTO ZERO)
Data stonage 500 data can be memorised
Data output RS232C Serial Interface

DIMENSIONS

dn500n-t.jpg

DN500N

  • Platform scale, Truck scale,
  • Tension/compression tester, Horse weigher,etc.
  • 50 kind of parts number can be memorized
  • Watch-dog function installed for diagnostics
  • Data back up against power stoppage.
  • Front display window covered by lexan film for the protecting of dust
  • Weighing judgment by over & under setting when printer is connected
  • Three kinds of hold function (peak-hold and averagehold, auto-hold)
  • Can drive up to 8pcs of load cell (350.)
  • RS-232C and CURRENT LOOP standard installed

SPECIFICATIONS

Specifications Accuracy
Load cell excitation 10VDC±5%, 320mA
Input Signal 0.5 ~ 3.0mV/V
A/D converting speed 200times/sec
A/D converter resolution 1/520,000 counts(19bits), Δ∑
Non-linearity 0.01% of full scale
Zero adjustment range -0.6mV~42mV
Display Unit 7 segment, 20.0mm(H)×13.0(W)mm
Annunciator steady, zero, tare, gross, auto, print, hold,Interface
Decimal point position 0, 0.0, 0.00, 0.000
Keyboard Number key and function key 16 PCS.
Temperature-drift zero ±20PPM/℃, span ±20PPM/℃
Non-linearity 0.01% of full scale
Rear panel Loadcell input : 5pin connector
External control : 12pin input terminal
(zero, tare,tare reset/hold/print, gross/hold reset/subprint,Current loop, RS232C)
Power SMPS Free Voltage (85V~265V)

Option

  • Analog Out-put (V-out) : 0~10V
  • Analog Out-put (I-out) : 4~20mA
  • Serial I / F : RS-422, RS-485
  • BCD In-put
  • BCD In-out
  • RS3232C and Current loop : Standard Installed

DIMENSIONS

สินค้าที่คล้ายกัน

B 10 Concentricity Gage

B-10 Universal Punch Concentricity Gage

Model B-10 Concentricity Gage

Model Description: Basic Extended Gage /
Catalogue En 2017 Page 01

เครื่องคัดแยกอัตโนมัติ ตรวจสอบแม่นยำด้วย กล้องตรวจสอบชิ้นงานขนาดเล็ก จากประเทศไต้หวัน

เครื่องตรวจสอบและคัดแยกชิ้นงานด้วยกล้องอัตโนมัติ เจ้าของธุรกิจขนาดเล็กเเละขนาดใหญ่ทุกท่าน อุสาหกรรมเกี่ยวกับน็อต สกรู จะรู้จักเครื่องประเภทนี้เป็นอย่างดี เเละมีการนำเข้ามาจำหน่ายกันหลากหลายยี่ห้อ เเล้วทำไมต้องเป็น (Accuvision)
Kelch Precision

KELCH – Precision and dynamics

KELCH – Precision and dynamics for your manufacturing กว่า 70

PRODUCT INFORMATION

Buyer information